IMPORTANT NOTICE:
• When used correctly, the device does not come into contact with the lesion nor the skin.
• When used normally, there is no need to perform any maintenance, cleaning or disinfection.
• Never clean the thread of frontend and backend. The existing coating should be protected.
• Should the applicator touch the patient in case of misuse, the applicator can be cleaned (autoclaved: 134 °C/273 °F - 4 min
or 121 °C/250 °F – 20 min) and disinfected (use a non-corrosive disinfecting solution, follow the disinfecting solution
manufacturer's recommendations with regards to the concentration and time of application).
• Always gently manipulate all components of CryoPen instruments. Never use more than gentle hand force. Never use pliers
or other tools to manipulate the instrument.
• Please ensure to only use the cartridges and filter/openers supplied from H&O Equipments nv/sa. Any other cartridges and
filter/openers would damage the instrument.
• Incorrect placement of the filter/opener device into the loading cavity will cause irreversible damage to the instrument.
• Residual foreign matters in the loading cavity while placing a new filter/opener can cause instrument failure.
• When loaded, the instrument is under high pressure. Any change to the construction of the instrument may cause a potential risk.
• Never put your hand on the gas release opening at the end of the backend while removing the cartridge!
• All external parts can be wiped with a cloth soaked in any noncorrosive disinfecting solution. Never immerse the entire
instrument in a disinfecting solution.
• The applicators must be kept away from dust and the cartridges must be kept away from high temperatures (max
50°C/122°F). Leave the applicator cap on when not in use.
• The CryoPen should be handled with care. When storing the instrument with a loaded cartridge in its case, position the CryoPen
in such a way that the trigger cannot be pressed when closing the lid of the case. This would cause undesired gas flow.
• CryoPen instruments are intended for professional use only.
• The N
O cartridges contain both a liquid and a gaseous phase in balance. The proportion of each phase depends on the room
2
temperature which affects the output flow. The best performances are achieved when the cartridge temperature is between
19°C and 22°C (66°F and 72°F). Lower and higher temperatures will give different values of pressure and density of the
cryogen. Over 25°C/77°F, the quantity of liquefied N
• Empty cartridges should be discarded as metal scrap.
O in the CryoPen jet will decrease.
2
7