ELECTROMAGNETIC COMPATIBILITY
The units with PROCOM2 or C2020 automate needs special precautions regarding EMC.
The units with PROCOM2 or C2020 automate has to be installed and put into service according to the EMC requirements joined in the following tables
Guidance and manufacturer's declaration – electromagnetic emissions – for all ME equipment and ME systems
The units with PROCOM2 or C2020 automate is intended for use in the electromagnetic environment specified below. The customer or the user of the units with
PROCOM2 or C2020 automate should assure that it is used in such an electromagnetic environment.
Emissions test
RF emissions
CISPR 11
RF emissions
CISPR 11
Harmonic emissions
IEC 61000-3-2
Voltage fluctuations/
flicker emissions
IEC 61000-3-3
Guidance and manufacturer's declaration – electromagnetic immunity - for all ME equipment and ME systems
The units with PROCOM2 or C2020 automate is intended for use in the electromagnetic environment specified below. The customer or the user of the units with
PROCOM2 or C2020 automate should assure that it is used in such an electromagnetic environment.
Immunity test
IEC 60601 test level
Electrostatic
discharge (ESD)
IEC 61000-4-2
Electrical fast
transient/burst
±2 kV for power supply lines
IEC 61000-4-4
±1 kV for input/output lines
Surge
±1 kV line(s) to line(s)
IEC 61000-4-5
±2 kV line(s) to earth
<5 % UT (>95 % dip in UT) for 0,5
cycle
Voltage dips, short
40 % UT (60 % dip in UT) for 5
interruptions and voltage
cycles
variations on power
70 % UT (30 % dip in UT) for 25
supply input lines
cycles
IEC 61000-4-11
<5 % UT (>95 % dip in UT) for 5
Power frequency (50/60
Hz)
magnetic field
IEC 61000-4-8
NOTE: UT is the a.c. mains voltage prior to application of the test level.
42
Compliance
Group 1
Class B
Class A
Complies
Compliance level
For 230 Vac / 50Hz
±6 kV contact
±6 kV contact
±8 kV air
±8 kV air
For 230 Vac / 50Hz
2 kV for power supply lines
±1 kV for input/output lines
For 230 Vac / 50Hz
±1 kV mode differential
±2 kV mode common
Pour 230 Vac / 50 Hz
0 Vac during 10 ms
92 Vac during 100 ms
161 Vac during 1 s
0 Vac during 5 s
s
3 A/m
3 A/m for 50Hz
Electromagnetic environment – guidance
The units with PROCOM2or C2020 automate uses RF energy only for its internal
function. Therefore, its RF emissions are very low and are not likely to cause any
interference in nearby electronic equipment.
The units with PROCOM2 or C2020 automate is suitable for use in all establishments,
including domestic establishments and those directly connected to the public low
voltage power supply network that supplies buildings used for domestic purposes
Electromagnetic environment – guidance
Floors should be wood, concrete or ceramic tile. If floors are
covered with synthetic material, the relative humidity should
be at least 30 %.
Mains power quality should be that of a typical commercial or
hospital environment.
Mains power quality should be that of a typical commercial or
hospital environment.
Mains power quality should be that of a typical commercial or
hospital environment. If the user of the units with PROCOM2
or C2020 automate requires continued operation during
power mains interruptions, it is recommended that the
emergency mobile system for air, O2 and vacuum networks
with PROCOM2 automate be powered from an uninterruptible
power supply or a battery.
Power frequency magnetic fields should be at levels
characteristic of a typical location in a typical commercial or
hospital environment.
No components sensible in product to magnetic field.
520713-00