Spare Parts And Their Replacement; Replacing Fi Lters; Battery Charging; Battery Replacement - cleanAIR MedicAER Manual Del Usuario

Tabla de contenido

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

  • MEXICANO, página 20
6.

Spare parts and their replacement

6.1.

Replacing fi lters

Only fi lters designed for CleanAIR® MedicAER® may be used! The fi lters must be new, unused and in the original pack-
aging. The expiration date must not have elapsed – it is marked on a sticker on the body of the fi lter.
Filter replacement:
1. Unscrew the fi lter from the unit.
2. Check the rubber sealing ring at the inputs to the unit for possible damage.
3. Fasten the new fi lters onto the unit and screw them tight.
There are two fi lter threads RD40x1/7" on the unit (standardized according to EN 148-1). Whenever changing fi lters,
should always change a set of two fi lters of the same type at the same time!
It is possible to use a pre-fi lter. This catches larger particles and especially aerosols, which can clog up the fi lter during
paint-spraying. Thus a pre-fi lter prolongs the life-time of the fi lter.
Install only new fi lters without any visible damage.
It is forbidden to clean a fi lter or blow through it in any way!
From the hygiene point of view it is not recommended to use fi lters for more than
6.2.
Batteries
For technical reasons, batteries are not fully charged when leaving the manufacturer. They are supplied with a 50 -
70% charge. Therefore either fully charge the battery before fi rst use, or expect shorter runtime in the fi rst cycle. For
maximum battery life, follow the storage instructions (see Section 9. "Storage")

6.2.1. Battery charging

Slide the battery out of the unit.
Connect the charger to the 100 - 230 V ~ 50/60 Hz mains.
Connect the battery to the charger. Charging is indicated by a red light, the green light indicates connection to
power. Charging time is < 3 hours.
When the red light goes off , the battery is fully charged.
After charging, disconnect the battery from the charger and the charger from the mains.
It is also possible to charge the battery while the battery is still plugged into the unit. The unit must be turned off
during charging.
It is not recommended to leave the charger connected to the mains when not in use.
Charging starts after connecting the battery using the charger to the power supply. After the battery is fully charged,
the charger switches to maintenance mode to keep the battery fully charged. The charging time is less than 3 hours.
Red LED on: fast charging
Red LED fl ashing while plugged: error
Red LED off : charge complete
Red LED fl ashing while unplugged: pre-charge detection
The battery charger is for indoor use only. Never charge the battery in potentially explosive atmosphere.
It is forbidden to use the battery charger for other purposes than those determined by the producer.
6
you
1 month after their fi rst use.
6.2.2.

Battery replacement

see pictorial annex "How to insert the battery" and "How to remove the battery"
1. Grip the powered respiratory unit by the rim on the top side of the battery with your left hand. Release the safety
lock located on the bottom of the unit between the fi lters. Then push the battery out of the unit using your thumb
(press the rubber place on the battery).
2. Insert the battery into the corresponding opening of the unit until the click-in stop locks the battery in the right
position.
7.
Possible faults and alarms
In case of any fault, such as a sudden decrease or increase in air supply when the user is in a contaminated working
area, it is necessary to leave the worksite immediately.
If there is a problem, check the following:
Whether the unit is assembled correctly.
Condition of the battery.
Filter clogging.
Whether the sealing of the safety hood provides a good fi t.
The unit is equipped with warning systems that are activated in the following situations:
Blocked fi lter - if the unit cannot main-
Low battery charge - an alarm
tain the chosen airfl ow, the electron-
notifi es the user when the
ics will automatically decrease the
battery has run out. It is best
airfl ow by one level and you will hear
to leave the contaminated
an acoustic signal.
If the unit is unable
area and replace or recharge
to maintain the lowest possible airfl ow,
the battery.
you will be warned by an audiovisual
alarm.
Filter expiration – the alarm should be
pre-set manually before you start us-
ing new fi lters. Once you start using
the fi lters, the unit starts the count-
down from the pre-set time. When the
pre-set time expires, the unit will start
to warn you.
7