OHAUS a-AB33M1 Manual De Instrucciones página 26

Ocultar thumbs Ver también para a-AB33M1:
Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

EN-24
Fast
Medium
Slow (a-AB41PH)
Each selection's specific algorithm is as follows:
Selection
Smart stability with 0.1
resolution or 1 significant digit
Smart stability with 0.01
resolution or 2 significant digit
Smart stability with 0.001
resolution or 3 significant digit
Fast stability
Medium stability
Slow stability (a-AB41PH)
Note: please enter the following PH resolution menu to set resolution mentioned in the
first three selections in the table.
Average
Enable or disable average.
Auto Smart
OFF
PH (a-AB33M1, a-AB33PH, a-AB41PH)
Resolution
0.1
0.01
0.001 (a-AB41PH )
Note: please refer to Select a pH resolution section for details.
Buffer group
USA
= faster stabilization time with less stability.
= normal stabilization time with normal stability.
= slower stabilization time with more stability.
Table 8-2 Stability algorithm
When the difference between the maximum value
and the minimum value of the measurement result
within 4 seconds is less than 0.5pH or 0.6mV, the
meter/analyzer will consider the result as stable.
When the difference between the maximum value
and the minimum value of the measurement result
within 6 seconds is less than 0.08pH or 0.15mV, the
meter/analyzer will consider the result as stable.
When the difference between the maximum value
and the minimum value of the measurement result
within 8 seconds is less than 0.01pH or 0.1mV, the
meter/analyzer will consider the result as stable.
When the difference between the maximum value
and the minimum value of the measurement result
within 4 seconds is less than 0.5pH or 0.6mV, the
meter/analyzer will consider the result as stable.
When the difference between the maximum value
and the minimum value of the measurement result
within 6 seconds is less than 0.08pH or 0.15mV, the
meter/analyzer will consider the result as stable.
When the difference between the maximum value
and the minimum value of the measurement result
within 8 seconds is less than 0.01pH or 0.1mV, the
meter/analyzer will consider the result as stable.
= the meter/analyzer calculates the average automatically. This
selection helps to reach measurement stability faster.
= disabled.
Analyzer and Bench Meter
Algorithm

Publicidad

Tabla de contenido
loading

Este manual también es adecuado para:

A-ab33phA-ab41phA-ab33ec

Tabla de contenido