EN-26
Enter temperature value (Enter temp.value)
5.0~60.0℃ (41.0~140℉) (the default is 25.0℃/77.0℉)
Note: please refer to Set temperature unit and temperature for details of the
temperature setting.
Temperature calibration (
Enter this menu to set calibration temperature if the temperature measured by your
electrode is incorrect or you change to a different temperature probe to use.
NIST-traceable thermometers should be used to measure and verify the temperature to
enter. Please carefully inspect the certified thermometers to ensure that they conform to
the specifications and are certified against NlST standards.
For how to use NIST-traceable thermometers to measure the temperature and input the
temperature to the meter:
1.
Put the electrode connected to the meter to a known and temperature stable
solution.
2.
Put the NIST thermometer to the same solution to measure temperature.
3.
Wait for the meter's temperature to become stable and check against the
temperature measured by the NIST thermometer.
4.
If you need to enter the temperature measured by the NIST thermometer to the
meter, touch the Menu icon
5.
Touch Setup, and then touch Temperature.
5.
Touch Temp.cal. Press the
value on the screen to confirm your setting.
The available temperature value is: 0~100℃ (32~212℉)
8.2
i-Steward menu
Enter this menu to set electrode condition parameters. The default settings are in bold.
Menu
Sub-Menu
Cal.Due Alarm
PH ECS
Cal log list (a-AB41PH)
Self-Test
i-Steward
Log Clear (a-AB41PH)
Reset (a-AB41PH)
Calibration due alarm (Cal.Due Alarm)
Enter this menu to enable or disable calibration alarm.
OFF
1 day
Temp.cal)
to enter menu option.
or
Table 8-3 i-Steward menu
= disabled
= the meter/analyzer will warn if no calibration within 1 day after
the last calibration.
Analyzer and Bench Meter
button to adjust the value. Touch the
Options
OFF, 1 day, 7 days, 14 days
/
/
/
/
Factory reset
Parameter reset
Reset password