Descargar Imprimir esta página

Reprocessing And Maintenance - Stryker 1488 Manual Del Usuario

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

  • MEXICANO, página 155

Reprocessing and Maintenance

These reprocessing instructions are provided in accordance with ISO 17664,
AAMI TIR12, AAMI ST79, and AAMI ST81. While they have been validated by
Stryker as being capable of preparing the device for re-use, it remains the
responsibility of the processor to ensure that the reprocessing as actually
performed, using equipment, materials, and personnel in the reprocessing
facility, achieves the desired result. This normally requires validation and
routine monitoring of the process. Stryker recommends users observe these
standards when reprocessing medical devices.
Warnings
• This device must be cleaned and sterilized prior to the first use and
after every subsequent use.
• Use only the sterilization cycles outlined in this document. Using
unspecified sterilization cycles may damage the device or result in
incomplete sterilization.
• Separate the camera and scope prior to cleaning, disinfection, or
sterilization.
• Wear appropriate protective equipment: gloves, eye protection, etc.
Cautions
• Always install the soaking cap prior to processing the camera. Failure
to properly tighten the soaking cap will corrode the connector pins
and void the warranty.
• Inspect the camera cable for cuts and breaks before soaking in any
fluid. Return any damaged camera to Stryker for service.
• Never soak the camera in the same tray with sharp instruments.
• Do not use brushes or pads with metal or abrasive tips during
manual cleaning, as permanent scoring or damage could result.
• To minimize galvanic corrosion, avoid soaking dissimilar metals in
close proximity.
• The Pendulum Camera is not autoclavable. Steam sterilizing camera
heads that are not marked
• Allow the camera head to cool before connecting it to the console.
Connecting the camera head while it is hot may result in system
error.
• When using Steris
head from the chamber once sterilization is complete, or moisture
may condense inside the camera head and cause display defects.
autoclave
®
liquid chemical sterilization, remove the camera
will result in product damage.
EN-15

Publicidad

loading

Este manual también es adecuado para:

1488310130