Steris System 1 / 1E / 1 Plus / 1 Express
Note:
Steris System 1®, System 1® Plus, and System 1® Express are
not intended for use in the United States.
1.
Clean and prepare the camera head and cable as recommended in this
user guide. Ensure the soaking cap is installed.
2.
Following the instructions of the manufacturer, sterilize the camera
head and cable using one of the Steris systems below with the
appropriate sterilant:
• System 1 with Steris 20 Sterilant
• System 1E® with S40™ Steris Sterilant
• System 1 Plus with S40 Sterilant
• System 1 Express with S40 Sterilant
3.
Remove the camera head and cable from the Steris chamber once
sterilization is complete, or moisture may condense inside the camera
head and cause display defects.
4.
Allow the camera head, cable, coupler, and scope to completely dry
before reassembly. Any moisture on the threads will cause the camera
and coupler windows to fog during use.
Steris/Amsco V-PRO
1.
Clean and prepare the camera head and cable as recommended in this
user guide. Ensure the soaking cap is installed.
2.
If using a sterilization tray (optional), follow any additional instructions
provided with the tray. Use only trays that are approved for sterilization
with Steris/Amsco® V-PRO®.
3.
Double wrap the camera head and cable (or tray) prior to sterilization.
4.
Sterilize the device using the V-PRO maX Sterilizer (Non-Lumen or
Standard cycle), the V-PRO 1 Plus Sterilizer (Non-Lumen or Standard
cycle), or the V-PRO 1 Sterilizer (Standard cycle).
5.
Allow the camera head, cable, coupler, and scope to completely dry
before reassembly. Any moisture on the threads will cause the camera
and coupler windows to fog during use.
EN-21