Descargar Imprimir esta página

Stryker 1488 Manual Del Usuario página 22

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

  • MEXICANO, página 155
5. Rinse
• Thoroughly rinse the device with RO/DI water until all detergent
residue is removed. Flush any lumens or crevices a minimum of five
times. Once all detergent residue is removed, continue to rinse for a
minimum of 30 seconds.
• Drain the excess water from the device and dry it using a clean cloth
or pressurized air.
Prolystica® 2x Neutral Detergent is validated for cleaning efficacy.
3
Cleaning: Automated
1. Brush
• Using a syringe, inject 50 mL of the enzymatic detergent (from
the "Preparation for Cleaning" section) into any lumen and mated
surface a minimum of one time.
• Brush from both ends of any lumens a minimum of five times, using
an appropriate bottle brush.
2. Rinse
• Rinse the device with RO/DI water at ambient temperature
until there is no visible detergent residue. Continue to rinse for
a minimum of 30 seconds after all detergent residue has been
removed.
• Place the device in the washer on an incline to facilitate drainage.
3. Automated wash
• Program the washer using the following parameters:
Phase
Pre Wash
Enzyme
Wash
Wash 1
Rinse 1
Dry Phase
EN-18
Recirculation
Time
2 minutes
2 minutes
2 minutes
2 minutes
7 minutes
Water
Temperature
Cold tap
water
Hot tap water
Set point
(66 ˚C)
Hot tap water
115 ˚C
Detergent Type and
Concentration
(if applicable)
N/A
Enzymatic Detergent
Non-enzymatic
Detergent
4
N/A
N/A

Publicidad

loading

Este manual también es adecuado para:

1488310130