Descargar Imprimir esta página

Electromagnetic Compatibility Information - VISIOMED TensioFlash KD-558 Manual Del Usuario

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

  • MEXICANO, página 24

ELECTROMAGNETIC COMPATIBILITY INFORMATION

Table 1
For all ME EQUIPMENT and ME SYSTEMS
Guidance and manufacture's declaration - electromagnetic emissions
The [KD-558] is intended for use in the electromagnetic environment specified below.
The customer or the user of the [KD-558] should assure that it is used in such an environment.
Emissions test
Compliance
RF emissions
Group 1
CISPR 11
RF emissions
Class B
CISPR 11
Harmonic emissions
Not applicable
IEC 61000-3-2
Voltage fluctuations/
flicker emissions
Not applicable
IEC 61000-3-3
Table 2
For all ME EQUIPMENT and ME SYSTEMS
Guidance and manufacturer's declaration - electromagnetic immunity
The [KD-558] is intended for use in the electromagnetic environment specified below. The
customer or the user of the [KD-558] should assure that it is used in such an environment.
IMMUNITY
IEC 60601test level
test
Electrostatic
± 6 kV contact
discharge (ESD)
± 8 kV air
IEC 61000-4-2
Power frequency
(50/60 Hz)
3 A/m
magnetic field
IEC 61000-4-8
156
Visiomed - KD-558 - 032016
Electromagnetic environment
- guidance
The [KD-558] uses RF energy only
for its internal function. Therefore,
its RF emissions are very low and are
not likely to cause any interference
in nearby electronic equipment.
The [KD-558] is suitable for use in
all establishments other than do-
mestic and those directly connected
to the public low-voltage power
supply network that supplies build-
ings used for domestic purposes.
Electromagnetic
Compliance level
environment - guidance
Floors should be wood,
concrete or ceramic tile.
± 6 kV contact
If floors are covered with
± 8 kV air
synthetic material, the
relative humidity should
be at least 30 %.
Power frequency magnetic
fields should be at levels
3 A/m
characteristic of a typical
location in a typical commer-
cial or hospital environment.
Table 3
For all ME EQUIPMENT and ME SYSTEMS that are not LIFE-SUPPORTING
Guidance and manufacturer's declaration - electromagnetic immunity
The [KD-558] is intended for use in the electromagnetic environment specified below. The
customer or the user of the [KD-558] should assure that it is used in such an environment.
IMMUNITY
IEC 60601test
test
level
RADIATED RF
3V/m 80 MHz
IEC 61000-4-3
to 2.5 GHz
NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is
affected by absorption and reflection from structures, objects and people.
a Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless)
telephones and land mobile radios, amateur radio, AM and FM radio broadcast and TV
broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic
environment due to fixed RF transmitters, an electromagnetic site survey should be considered.
If the measured field strength in the location in which the [KD-558] is used exceeds the
applicable RF compliance level above, the [KD-558] should be observed to verify normal
operation. If abnormal performance is observed, additional measures may be necessary, such
as re-orienting or relocating the [KD-558].
b Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3V/m.
Visiomed - KD-558 - 032016
Compliance
Electromagnetic environment
level
- guidance
3 V/m
Portable and mobile RF com-
munications equipment should
be used no closer to any part of
the [KD-558], including cables,
than the recommended separation
distance calculated from the equa-
tion applicable to the frequency of
the transmitter.
Recommended separation
distance:
d = 1.2 √ P
80 MHz to 800 MHz
d = 2.3 √ P
800 MHz to 2,5 GHz
Where P is the maximum output
power rating of the transmitter
in watts (W) according to the
transmitter manufacturer and d
is the recommended separation
distance in meters (m).
Field strengths from fixed RF
transmitters, as determined
by an electromagnetic site
survey,a should be less than
the compliance level in each
frequency range.b
Interference may occur in the
vicinity of equipment
marked with the following
symbol:
157

Publicidad

loading