Bath Circulator
Grounding Strap
and Nut
Thermo Scientific
The circulators is designed for continuous operation and for indoor use.
The equipment normally ships with the immersion circulator mounted facing
the reservoir. You may change the position ±90° by removing thumb screws,
no tools are required.
Do not mount it backwards; the line cord could contact the reservoir
fluid.
Rotate immersion
circulator only ±90°
from shown position
Never place the circulator in a location where excessive heat, moisture,
inadequate ventilation, or corrosive materials are present.
Leave refrigerated baths in an upright position at room temperature
(~25°C) for 24 hours before starting. This will ensure the lubrication oil
has drained back into the compressor.
Sahara baths, and stand-alone circulators, require immersion circulator
grounding.
For Sahara baths, and stand-alone immersion circulators, use the
supplied nut and washer to secure the grounding strap to the top of
the metal tank.
A45HC units utilize a flammable refrigerant, risk of fire or explosion,
do not use sharp objects to clean evaporator coil in the tank.
Section 3 Installation
3-21