Vertical recalibration
Place the inclination spirit level on a hard surface, which is
as vertical as possible, with the module at the top and look
at the LCD. Switch on the inclination gauge using the
Q button and wait 10 seconds. Then press the E button
for 2 seconds.
CAL 1 briefly appears in the display, and the measured
value then starts flashing. Turn the inclination spirit level
around its own axis so that the module is still at the top
and the horizontal vial is touching the surface. Wait 10
seconds, then press the E button again for 2 seconds.
CAL 2 briefly appears in the display.
The inclination gauge is now vertically calibrated.

Maintenance and care

The inclination spirit level is shockproof and waterproof.
You can clean it with a wet cloth but do not immerse it in water.
Calibration after the error message "CAL ALL"
1. Cal 1 and Cal 2 = CAL ALL
Place the measuring instrument on a horizontal surface
Let it rest for 15 seconds
Press the E button / Cal 1 then appears / the value starts flashing
Turn the measurement device around by 180°
Let it rest for 15 seconds
Press the E button / Cal 2 then appears / followed by CAL ALL
2. Cal 1 and Cal 2 = CAL ALL
Place the measuring instrument upside-down
(module pointing down)
Let it rest for 15 seconds
Press the E button / Cal 1 then appears / the value starts flashing
Turn the measurement device around by 180°
Let it rest for 15 seconds
Press the E button / Cal 2 then appears / followed by CAL ALL