Hultafors ELWX 60 Instrucciones De Uso página 19

Tabla de contenido

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

  • MEXICANO, página 39
3. Cal 1 and Cal 2
Place the measuring instrument against a vertical surface
with the vial at the top
Let it rest for 15 seconds
Press the E button / Cal 1 then appears / the value starts flashing
Turn the measurement device around by 180°
Let it rest for 15 seconds
Press the E button / Cal 2 then appears / followed by CAL ALL
4. Cal 1 and Cal 2
Place the measuring instrument against a vertical surface with
the vial at the bottom
Let it rest for 15 seconds
Press the E button / Cal 1 then appears / the value starts flashing
Turn the measurement device around by 180°
Let it rest for 15 seconds
Press the E button / Cal 2 then appears / followed by the value.
THE TOTAL CALIBRATION "CAL ALL" IS THUS COMPLETED
= CAL ALL
XX
,xx°
=
19

Publicidad

Tabla de contenido
loading

Productos relacionados para Hultafors ELWX 60

Este manual también es adecuado para:

Lwx 60Dm 60

Tabla de contenido