NewLife
®
Elite
IV. 11. Troubleshooting
Observations
The 0-I button is in the ON
position. The device does not
operate. The continuous alarm
sounds.
Oxygen concentration indicator
remains lit yellow.
The alarm test does not work.
The 0-I button is ON and the
compressor is operating but
there is no flow. The audible
alarm sounds.
The 0-I button is ON, the
compressor is operating, there
is a flow but the audible alarm
sounds.
The compressor stops in mid-
cycle, then starts again after a
few minutes.
The oxygen enriched air flow is
interrupted at the nasal
cannula outlet.
The flow at the nasal cannula
outlet is irregular.
AirSep
®
Corporation
Probable causes
Power cable not plugged in
Check the cable connection.
correctly. Mains power failure.
Reset the circuit breaker (2) if
necessary by pressing. Check
the fuses or circuit breaker
fitted on the premises.
Oxygen concentration is too
Contact your distributor.
low.
Internal electrical fault.
Contact your distributor.
Pneumatic connection broken
Stop the device by pressing the
or compressor problem.
0-I button and contact your
distributor.
Internal electrical fault.
Stop the device and contact
Pneumatic circuit fault.
your distributor.
Compressor thermal safety
Stop the device and wait for it
device has been tripped. Fan
to cool down. Check that the
not working.
patient circuit is not obstructed.
Start up again. If the device
does not start, contact your
distributor.
Tube disconnected or humidifier
Check the gas administration
not tight.
circuit.
Pneumatic circuit problem.
Contact your distributor.
GB
Solutions
9-GB