Cleaning; What To Do If - Seca 376 Manual De Instrucciones Y Garantía

Ocultar thumbs Ver también para 376:
Tabla de contenido

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

  • MEXICANO, página 124

8. CLEANING

9. WHAT TO DO IF ...?

Malfunction
... no weight is dis-
played during weighing?
0. 0 00
...
does not appear
before weighing?
... a segment lights up
continuously or not at
all?
...
is displayed?
bAtt
...
is displayed?
St0p
...
is displayed?
...
is displayed?
56 •
NOTE:
For further operation of the wireless printer see
its user manual.
To clean the tray and housing of the scales use a
household detergent or standard disinfectant as
required. Take note of the instructions provided by the
manufacturer.
Cause/Remedy
The scales are not connected to the power supply
- Check the scales are switched on
- Check batteries are inserted
A load was placed on the scales before being
switched on.
- Remove load on scales
- Switch scales off and back on again
There is a fault at that point.
- Notify maintenance service.
Battery voltage is falling off.
- Insert new batteries as soon as possible
Batteries are flat.
- Insert new batteries
The maximum load has been exceeded.
- Reduce load
The ambient temperature of the scales is too high
or too low.
- Set up scales in an ambient temperature
between +10 °C and +40 °C
- Wait approx. 15 minutes until scales have
adjusted to ambient temperature

Publicidad

Tabla de contenido
loading

Tabla de contenido