Maximum Quantity of Devices
WARNING: DO NOT exceed the specified maximum quantity
of devices in an insert tray. The maximum quantity of devices
is based on Stryker validation of this equipment. Exceeding
the maximum quantity of devices may prevent proper
sterilization of the equipment and/or result in excessive
weight. Always follow the current local recommendations and/
or regulations governing weights and weight limits.
Sterilization Case, Large (Four Handpiece) (REF
6102-450-000)
DEVICE TYPE
Handpieces
Attachments
Total of All Devices
Mini Sterilization Case (Two Handpiece) (REF
6102-452-000)
DEVICE TYPE
Handpieces
Attachments
Total of All Devices
Sterilization Case (Three Handpiece) (REF 6102-
454-000)
DEVICE TYPE
Handpieces
Attachments
Total of All Devices
Recommended Placement of Devices
WARNINGS:
• ALWAYS make sure the equipment is clean and dry before
sterilization.
• ALWAYS make sure handpiece and attachment lumens remain at
a vertical angle to allow water to sufficiently drain.
NOTES:
• See the Maximum Quantity of Devices section.
• The illustrated placement of devices is a recommendation only.
Other placement combinations are acceptable.
4
MAXIMUM QUANTITY
4
12
16
MAXIMUM QUANTITY
2
12
14
MAXIMUM QUANTITY
3
10
13
Sterilization Case, Large (Four Handpiece) (REF
6102-450-000)
Place the handpieces and attachments into the insert tray as shown.
Mini Sterilization Case (Two Handpiece) (REF
6102-452-000)
Place the handpieces and attachments into the insert tray as shown.
Sterilization Case (Three Handpiece) (REF 6102-
454-000)
Place the handpieces and attachments into the insert tray as shown.