VII. Operating Method

1. Boot Up

Before booting up, the trigger should be pressed for a couple of times to ensure the jaws are well
closed.
Press POWER key, and it is switched into the boot-up state. First to automatically test
LCD display, all its symbols show up, see Figure 1. Then to start the auto inspection; in
this process, it will be followed by showing "CAL6, CAL5, CAL4:CAL0, OL
Figure 2. When "OL " appears, auto inspection is completed, and then automatically
enter the resistance measurement model, see figure 3.
In the process of auto inspection, DO NOT press the trigger, nor open the jaw, nor
clamp any wire.
In auto-inspection process, be sure to maintain the natural static state of the Meter; do not
overturn the Meter, nor impose any external force on the jaw. Otherwise, the accuracy of
measurement cannot be guaranteed.
In auto-inspection process, if the jaws clamped around a conductor loop, the
measurement is not accurate. Please remove conductor loop and reboot it up.
If there was not an OL appearing after auto-inspection, but a greater resistance value
displayed, as shown in figure 4; But the test loop detection can still give out the correct
result. This shows that the Meter has a larger error only in measuring the major resistance
(e.g. more than 100 ), whereas in measuring the small resistance, it can still maintain the
original accuracy, users can be rest assured in use.