OPERATING INSTRUCTIONS (CONT)
5.
Application of the Foot sleeve(s).
a) Foot sleeves come in identical pairs and may be used on either foot.
b) Place the sleeve on the patient's foot ensuring the
hook and loop attachments are visible.
c)
Ensure a snug fit by placing two fingers around the back of
the sleeve to ensure the sleeve is not too tight. Routinely
inspect the patient's feet once compression commences.
d) Using the same procedures, apply the second model sleeve if needed.
e) If only one sleeve is to be used simply leave the unused
air outlet on the pump free (no sleeve attached).
6. Air hose(s) are required to connect the sleeve(s) to the pump
and are provided separately. Attach the sleeve(s) to the
air hose(s) using the connectors. Each tube has a male end
connector at one end and a female connector at the other.
The female end will fit to the male end on the sleeve.
7.
Attach the other end(s) of the air hose(s) (male end) to the
female connector(s)on the pump. If you need to disconnect the
tubing, press the tab on female connector and pull apart.
8. Press and hold the power button to turn pump on when
air hose(s) and sleeve(s)are properly connected.
9.
The pump will automatically detect single/dual and leg (calf/
thigh)/foot model of sleeve(s) during the initialization process
lasting approximately 10 seconds. Highlighted graphics will
display how many foot versus leg sleeves are attached.
10. After initialization, check to confirm the screen
graphic(s) displays the desired settings.
11. To activate the timer, turn the pump on and timing will commence.
The timer will reflect minutes of compression. When compression is
interrupted and the pump is turned off, the timer will suspend timing
until the patient is reconnected to therapy and therapy is continued.
12. To reset the timer press "TIMER RESET" for
5 seconds and it will reset itself.
13. To turn pump off, hold POWER button for 1 second.
14. To disconnect device remove plug from electrical outlet.
15. To remove the sleeves, pull the connectors
apart.