Contraindications & Cautions; Cleaning; Warning Troubleshooting Guide - Medline HEMO-FORCE MDS600INT2 Instrucciones De Operación

Ocultar thumbs Ver también para HEMO-FORCE MDS600INT2:
Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

WARNING TROUBLESHOOTING GUIDE

Indicates the error information for Hemo-Force®,
these include: Lo, Hi, GE and F.
Alarm
Fault Type
Codes
LO
Low Pressure
HI
High Pressure
GE
Sleeve Error
F
System Failure
Description
• Leg is less than
40mmHg pressure; foot
pressure is less than
120mmHg after
2 consecutive cycles.
• Air hose disconnected
mid-operation.
• Air hose is kinked.
• Leg is more than
40mmHg pressure; foot
is more than 120mmHg
after 2 consecutive cycles.
• No sleeves attached dur-
ing start up.
• Sleeves are not fully de-
flated during each restart.
• The initialization failed
due to a possible blockage
in air hose/sleeve con-
nectors.
• Pressure is greater
than set pressure after
two inflations.
• Electrical malfunction.
13
Troubleshooting
• Check sleeve application
to ensure proper fit.
• Check for leaks in the
sleeves or air hose set and
replace if faulty.
• Turn the pump off and
restart.
• Refer to a service profes-
sional if problem persists.
• Check air hose set for
any kinks.
• Refer to a service profes-
sional if problem persists.
• Attach sleeves to pump
and restart.
• Ensure sleeves are fully
deflated before restart.
• Check connectors for any
blockage and restart.
• Refer to a professional for
service if problem persists.
• Check air hose connection.
• Turn the pump off and
restart.
• Refer to a service profes-
sional if problem persists.

Publicidad

Tabla de contenido
loading

Tabla de contenido