Dimensions (W x D x H)
Enclosure
Weight
Pollution degree
Installation category
Protection class
Power requirements
Altitude
Operating temperature
Humidity
Sample temperature
Sample flow rate
Sample pressure
Temperature accuracy
UV lamps
User interface
TOC
Outputs
on page 14
Details
59.7 x 22.9 x 25.4 cm (23.5 x 9 x 10 in.)
IP46
13.6 kg (30 lb)
2
II
I
100/230 VAC, ±10%, 65 W, 50/60 Hz
4000 m (13,125 ft) maximum
10 to 40 °C (50 to 104 °F)
5-95%, non-condensing
1 to 95 °C (34 to 203 °F)
60 to 300 mL/min
0.70 to 7.0 kg/cm² (10 to 100 psi)
±0.5 °C
2x, with UV Detect technology
Color touch screen
Range: 0.5 to 2000 ppb as carbon
Accuracy: ±1 ppb or ±5%, whichever is greater
Repeatability: ±0.3 ppb or ±1%, whichever is greater
Detection limit: 0.5 ppb
Range for conductivity mode: 0.05 to 150 μS/cm at 25 °C
Accuracy: ±2% over the full range
Range for TOC mode: 0.05-5.0 µS/cm
Analog: Three isolated
Discrete: Four isolated, configurable