The syringe outlet tube has a blue ring on it. If anti-diffusion tips are necessary, remove the pre-
installed outlet tube from the syringe and install the tube from the application kit with the pre-installed
anti-diffusion tip.
Push the outlet tubes into the tube holder slots so that they are correctly attached.
5.8 Install the sensor
5.8.1 Install the legacy adapter
Go to
Connect the sensor
on page 13 if no legacy adapter is included in the application kit.
1. Connect the measuring, reference and temperature sensors to the legacy adapter.
2. Make sure that the instrument display shows the Home screen. Connect the legacy adapter cable
to a sensor socket on the rear panel of the instrument.