1. From the home screen, select the measurement application, then push Select. Application
information shows on the display.
2. Read the related "Application Note" from the USB applications key for more instructions.
3. If necessary, select an icon for more information or to change some data.
Option
Description
Information Shows more information about the application.
Operator
Changes the operator ID. Select from a list of applicable operators.
Sample
Sample Name: Change the specified name of the sample.
Type: Push the left and right arrow keys and select the sample type (Sample, QC
or Define blank) to be used for the measurement. If Define blank has been
selected before, two more sample types are available (QC with blank and
Sample with blank).
4. Fill a beaker with the recommended sample amount that shows on the display. If necessary, add
more of the solvent specified in the application note until the sensor is correctly installed in the
sample.
5. Carefully put a magnetic stir bar into the beaker. Make sure there is no liquid spill.
6. Attach the beaker to the sensor holder.
7. Make sure that the icon at the bottom of the display
show on the display adjacent to this icon. Refer to
that the tubes and sensor are correctly aligned.
8. Push Start to start the measurement. Measurement data shows on the display.
9. If the default stirring speed needs to be adjusted, push the up and down arrow keys to increase or
decrease the speed.
Note: This adjustment only applies to the current operation. The standard default stirring speed for the
application is not changed.
10. Two options are available during the procedure:
Option Description
Stop
Aborts the measurement and no results are calculated. If selected during the
Replicate Sample option, then all data in the series is lost.
Skip
Stops the current operation and goes directly to the next step in the procedure. Results
are then calculated from measurement data available before Skip is selected. Results
can be less accurate using this option.
11. When the measurement is complete, push the arrow keys to see the different measurement
views.
12. Push Next for these options:
Option
Replicate
Sample
New Sample
13. Push Exit to go back to the home screen.
8.3 Syringe verification
Use the syringe verification option to examine the syringe accuracy based on the ISO 8655 standard.
22 English
Description
Use this option to start the same titration on the same sample. This is used to
study the repeatability by successively analyzing several parts of the same
sample. At the end of each measurement, a window shows the average
value, the standard deviation and the relative standard deviation.
Use this option to start the same titration on a new sample. No standard
deviation or relative standard deviation measurements will be done.
is highlighted. Do the instructions that
Connect the sensor
on page 13 to make sure