FAQ - English

For information on operating elements, see
document.
For chapter indications, refer to the complete instruction manual on the internet.
Error
LED B
fails to go on. • Control unit not
Unit operating, but
no temperature at
the tip.
The tip fails to reach
the desired tempera-
ture.
Possible cause
switched on.
• Power supply inter-
rupted.
• Plug not fully inserted
in the control unit.
• Faulty control unit.
• Faulty transformer.
• Short circuit in the tip.
• Tip not properly
screwed in.
• Tip / handpiece contact
contaminated.
• Faulty tip.
• Handpiece cable not
properly plugged into
the control unit.
• Plug on handpiece
cable shows signs of
corrosion.
• Incorrect power supply
to the tip because the
OEM mains power
supply unit is not being
used.
Fig. 1
and

Corrective action

• Switch the control unit on at switch "A".
• Check the power supply and reconnect, as
required.
• Push the plug in until it unmistakably locks
in place.
• Have unit replaced by Service Department.
• Replace with spare (refer to the spare parts
list).
• Clean the tip and handpiece
(use compressed air).
• Replace the tip.
• Faulty heater handle --> Replace with spare
(refer to the spare parts list).
• Hand tighten tip.
• Clean the contact (use compressed air).
• Replace the O-ring (refer to the spare parts
list).
• Replace with a new tip
(refer to Sec. 13 / "Accessories").
• Fully insert the handpiece cable into the
socket until the cable locks in place.
• Reinsert the plug, clean the contacts, if
required.
• Use the OEM mains power supply unit.