FAQ - English
For information on operating elements, see
For chapter indications, refer to the complete instruction manual on the internet.
Trouble
No display.
• Appliance not connected to mains.
• Appliance not turned on.
• No power from the socket.
• Faulty fuse.
Cannot change
• Appliance still running.
cleaning time.
Cannot save
• Cleaning time was entered during an
cleaning time.
interrupted cleaning process.
Insufficient clean-
• Cleaning time too short.
ing performance.
• Active ingredients of the cleaning fluid
spent.
• Insufficient number of cleaning needles.
• Cleaning needles spent.
Cleaning needles
• Cleaning needles have been insufficient-
have corroded.
ly rinsed.
• Cleaning needles have been stored
incorrectly.
• Cleaning needles in use for longer peri-
od of time.
All manuals and user guides at all-guides.com
Fig. 1
Possible Cause
- 1 -
on page 2 of this document.
Corrective Action
• Check correct fit of power cable in the appliance socket and
wall socket.
• Turn on appliance at the mains switch
• Check power socket, for example, with another appliance.
• Replace the fuse (see section 6.8). If the fuse blows repeat-
edly, have the appliance inspected by a service technician.
• Interrupt the cleaning process with the Start / Pause button,
and then adjust cleaning time.
• Stop the cleaning process with the Stop button, and then
enter new cleaning time.
• See chapt. 5.4.
• Prepare fresh cleaning fluid after each application.
• Fill up the cleaning needles (see section 6.5).
• Replace the cleaning needles (see section 6.6).
• Rinsing and storage see chapter 6.1.
• Removing the corrosion:
Clean the needles with 100 ml SYMPROfluid Universal,
without contents, for 5 minutes at 2000 rpm. Dispose of
cleaning liquid.
• The needles are socalled consumable goods. If it is not
possible to regenerate the needles, or the needles' perfor-
mance deteriorates, see chapter 6.6.
EN
(12, Fig.
1).