Operating altitude
Storage and transport temperature
Storage and transport humidity
Housing construction
Supplemental oxygen
Hypoallergenic air filter
Standard air filter
SlimLine
air tubing
TM
Standard air tubing
ClimateLine heated air tubing
MAX
ClimateLine
Air outlet
Electromagnetic compatibility
Aircraft use
IEC 60601-1 classification
8
English
heated air tubing
Sea level to 2,591 m
-20ºC to +60ºC
10–95% non-condensing
Flame retardant engineering thermoplastic
Recommended maximum supplemental oxygen flow:
4 L/min (VAuto or ASV mode); 15 L/min (S, ST or T mode)
Non-woven acrylic and polypropylene fibers with
polypropylene carrier
Polyester fiber
Flexible plastic, 1.8 m, 15 mm inner diameter
Flexible plastic, 2 m, 19 mm inner diameter
Flexible plastic and electrical components, 2 m, 15 mm
inner diameter
Flexible plastic and electrical components, 1.9 m, 19 mm
inner diameter
The 22 mm conical air outlet complies with ISO 5356-1
Product complies with all applicable electromagnetic
compatibility requirements (EMC) according to IEC60601-1-
2, for residential, commercial, and light industry
environments. Information regarding the electromagnetic
emissions and immunity of these ResMed devices can be
found on www.resmed.com, on the Products page under
Service and Support. Click on the PDF file for your language.
Medical-Portable Electronic Devices (M-PED) that meet the
Federal Aviation Administration (FAA) requirements of
RTCA/DO-160F can be used during all phases of air travel
without further testing or approval by the airline operator.
ResMed confirms that the S9 meets RTCA/DO-160F
requirements.
Class II (double insulation), Type BF