Operating Instructions; How To Modify Pump Settings - Medline Compat 9517455 Manual Del Usuario

Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

EN
4
indicates total volume delivered.
Dose Limit Display, Indicates Dose Limit
Volume set when Dose Limit Set Key is pressed
during operation.
Accumulated Volume Display, indicates
Accumulated Volume when Dose Limit Set Key is
pressed and held for more than five seconds.*
10
Rotary peristaltic pump mechanism
11
Power cord
12
Alarm volume control
Pole clamp
13
14
Directions
Cord wrap
15
Pump adapter bracket
16
Flow sensors and drip chamber guide
17
Tubing guide
18
*with serial numbers greater than 400000
Controls 11; 12; 13; 15 are on the back of the pump
Control 14 is on the side panel

Operating Instructions

(see Diagrams on page iv)
DIAGRAM 1:
Activate pump; set desired flow rate.
DIAGRAM 2:
Insert pump set drip chamber. Make sure set is
properly seated on pump.
DIAGRAM 3:
Assemble silicone tubing around rotor.
DIAGRAM 4:
Insert pump adapter into bracket.
DIAGRAM 5:
Place tubing into tubing guide.
Silicone Tubing that is not properly secured around
the pump's rotor may result in a free flow situation.
Always ensure that the pump adapter is correctly
seated into the bracket. This is particularly important
when using water or clear liquids, as proper alarm
function may not occur if the giving set is not
secured properly.
1. Plug the power cord into a 120 VAC., 60 Hz outlet.
The internal backup battery should normally only
be used for short term use when AC outlet power
is unavailable.
2. Press ON - alarm will sound briefly. Remember, the
pump has memory and will automatically recall
the previous parameter settings. (DIAGRAM 1)
3. Check Flow/Infusion Rate. If change is desired,
press INC or DEC to change rate setting.
(DIAGRAM 1)
4. Check Dose Limit by pressing DOSE LIMIT SET.
If change is desired, press INC or DEC within the
next 3 seconds to change setting limit. Zero Dose
Limit implies no dose limit monitoring by pump.
(DIAGRAM 1)
5. Press TOTAL RESET to clear Total Volume,
if desired.
6. Insert drip chamber into drip chamber brackets.
(DIAGRAM 2)
7. Open roller clamp. Expel air from tubing.
Close clamp.
8. Stretch silicone tubing around rotor. Be sure not
to overstretch the silicone tubing. Insert pump
adapter into bridge. (DIAGRAMS 3+4)
9. Place tubing into tubing guide. (DIAGRAM 5)
10. Attach pump set to patient's feeding tube.
11. Open roller clamp.
12. Press RUN/HOLD to start pump.
13. OFF will turn off the pump at any time.
(Flow/Infusion Rate, Dose Limit, Volume*
Delivered and Accumulated Volume are retained.)
*with serial numbers greater than 400000

How to Modify Pump Settings

To change rate or dose limit:
1. Press RUN/HOLD to stop pump.
2. Select new Flow/Infusion Rate by pressing INC
or DEC.
3. Select new Dose Limit by pressing
DOSE LIMIT SET and then press INC or DEC
within next 3 seconds.
4. Press TOTAL RESET to clear Volume Delivered,
if desired.
5. Press RUN/HOLD to restart pump.
To view dose limit set:
Press DOSE LIMIT SET. Dose Limit will be displayed
for a few seconds.