Alarms/Alerts - Medline Compat 9517455 Manual Del Usuario

Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

EN
2
with healthcare facility policy and local ordinances
for disposal of medical electrical equipment.
• This pump is intended for use only with the
available range of COMPAT
Standard
®
Administration Sets and only if the Administration
Set is correctly installed. Use with other
Administration Sets could cause delivery errors
and put the patient at risk.
• Caution: To reduce the risk of electrical shock,
do not remove back cover. Refer servicing to
qualified service personnel.
• Notice: To insure proper battery operation, unit
should be charged for 12 hours prior to use.
Remove label prior to use.
• Danger: Possible explosion hazard if used in the
presence of flammable anesthetics.
• The COMPAT
Enteral Feeding Pump is MEDICAL
®
ELECTRICAL EQUIPMENT requiring special
precautions regarding EMC and needs to be
installed and put into service according to the
EMC information in the ACCOMPANYING
DOCUMENTS.
• The COMPAT
Enteral Feeding Pump emits Radio
®
Frequencies (RF) that can affect other MEDICAL
ELECTRICAL EQUIPMENT located nearby.
ADMINISTRATION SETS
• Administration Set tubing may present a
risk of entanglement or strangulation. Arrange
giving set tubing to minimize the risk of patients or
others from strangulation or entanglement.
• Dispose of Administration Set according to
healthcare facility procedure for treatment of
medical waste, or as directed by a health care
provider.
• Follow instructions to reduce the risk of free flow.
• Only for enteral nutrition.

Alarms/Alerts

High Priority Alarm - Flashing red displays and
alarm sounds
Indicates an imminent hazard requiring immediate
operator attention, which otherwise could result in
serious personal injury and/or product damage if
proper action is not taken:
- FREE FLOW
Medium Priority Alarms - Flashing yellow indicator
light and alarm sounds
Indicates a condition that requires prompt operator
attention, which could result in serious personal injury
and/or product damage if proper action is not taken:
- OCCL/EMPTY
- LOW BATTERY
Alerts - Indicator light is lit and/or alert sounds
Indicates that a condition exists of which the
operator should be aware:
- LOW BATTERY (light)
- BAT CHARGING (light)
NOTE: Bat Charging light on serial numbers
starting at 850000
- RATE CHANGE (alert)
- DOSE COMPLETE (light and alert)
- BATTERY (light)
- HOLD (alert)
Cancel Alarm/Alert
Pressing the RUN/HOLD key after the alarm
condition has been corrected cancels the alarm/alert
and returns pump to hold mode.
NOTE: See Troubleshooting Section for possible
alarm/alert causes.
FREE FLOW ALARM CONFIRMATION TEST
To determine if the Free Flow Alarm has been
installed on the pump, complete the following steps
to determine if the pump has been upgraded to
include the Free Flow Alarm feature:
1. Turn pump on
2. Press and hold down both the INC and DEC
buttons at the same time until the pump beeps
3. Release the buttons
a. Below is the information that will be shown:
RATE Display
VOLUME DELIVERED Display
888
8888
4. Press and hold down both the INC and DEC
buttons at the same time until the pump beeps
5. Release the buttons
a. Below is the information that will be shown:
RATE Display
VOLUME DELIVERED Display
012
3456
6. Press and hold down both the INC and DEC
buttons at the same time until the pump beeps