ELECTROMAGNETIC IMMUNITY

The COMPAT
Standard Enteral Feeding Pump is intended for use in the electromagnetic environment specified below. The customer or the user of the COMPAT
®
Standard Enteral Feeding Pump should ensure that it is used in such an environment.
Guidance And Manufacturer's Declaration-Electromagnetic Immunity For All Equipment And Systems
Immunity Test
Electrostatic Discharge (ESD)
IEC 61000-4-2
Electrical fast transient/burst
IEC 61000-4-4
Surge
IEC 61000-4-5
Voltage dips, short interruptions
and voltage variations on power
supply input lines
IEC 61000-4-11
Power frequency (50/60 Hz)
magnetic field
IEC 61000-4-8
NOTE: UT is the a. c. mains voltage prior to application of the test level.
IEC 60601 Test Level
Compliance Level
± 6 kV (contact)
± 6 kV (contact)
± 8 kV (air)
± 8 kV (air)
± 2 kV for power supply lines
Not Applicable
± 1 kV for input/output lines
Not Applicable
± 1 kV differential mode
Not Applicable
± 2 kV common mode
Not Applicable
< 5% UT (>95% dip in UT ) for
Not Applicable
0,5 cycle
40% UT (60% dip in UT ) for 5
cycles
70% UT (30% dip in UT ) for 25
cycles
< 5% UT (>95% dip in UT ) for
5 sec
3 A/m
3 A/m
Electromagnetic Environment Guidance
Floors should be wood, concrete or ceramic tile. If floors
are covered with synthetic material, the relative humidity
should be at least 30%.
Mains power quality should be that of a typical commercial
or hospital environment.
Mains power quality should be that of a typical commercial
or hospital environment.
Mains power quality should be that of a typical commercial
or hospital environment. If the user of the COMPAT
Standard Enteral Feeding Pump requires continued
operation during power mains interruptions, it is
recommended that the COMPAT
Standard Enteral Feeding
®
Pump be powered from an uninterruptible power supply or
a battery.
Power frequency magnetic fields should be at levels
characteristic of a typical location in a typical commercial
or hospital environment.
®
®