To clear volume delivered:
1. Press RUN/HOLD to stop pump.
2. Press TOTAL RESET to clear Volume Delivered.
3. Press RUN/HOLD to restart pump.
To view or clear accumulated volume:
1. Press RUN/HOLD to stop pump.
2. Press and hold DOSE LIMIT SET for 5 seconds.
tot VOL and accumulated volume will be
displayed for 10 seconds.
3. To clear accumulated volume, press TOTAL
RESET within the 10 seconds.
4. Press RUN/HOLD to restart pump.
To view or change dose complete alarm mode:
Three alarm modes are available when dose is
complete:
• Mode 1 - Pump beeps then stops.
• Mode 2 - Pump beeps, stops and then beeps at 5
minute intervals.
• Mode 3 - Pump beeps continuously.
To select one of the desired modes:
1. Press INC and DEC for 5 seconds. The Infusion
Rate and Volume Delivered Displays should show
888 and 8888 respectively.
2. Press INC and DEC again. The Infusion Rate
Display will show 012 and the Volume Delivered
Display will show 3456.
3. Press INC and DEC once again. The Infusion Rate
Display will show the current alarm mode (1r, 2r
or 3r) (Disregard the Volume Delivered Display).
4. Select desired mode by pressing INC or DEC.
5. Press RUN/HOLD to save new alarm mode.
NOTE: If RUN/HOLD is not pressed, new alarm
mode will not be saved.
6. Turn pump off.
7. Pump is now set to operate in the new alarm mode.
Self Test
The major functions of the pump may be tested by
pressing the ON button and then simultaneously
holding the INC and DEC keys for three seconds.
The volume delivered and rate displays should show
all eights (8888) and the pump rotor should turn.
To test the alarms, press DOSE SET, INC, DEC,
TOTAL RESET or RUN/HOLD.
When any of these keys are pressed or the flow
sensors detect a drop, all alarm lights should light,
the alarm should sound and the rotor will halt
momentarily.
Any failure indicates a service check should be
performed. See Service Section for instructions on
reporting a pump in need of service.
Service
COMPAT
Enteral Feeding Pumps are reliable
®
medical electromechanical devices. As with any
mechanical device, problems may occur. The
following tips will help you correct the minor in-
service problems that may arise.
NOTE: Upon receipt of a new pump, plug it in
for a minimum of 12 hours to charge the backup
battery. For maximum battery life, charge the
battery at least every 30 days, even if the pump is
not in use. Whenever the backup battery is used
to power the pump, charge for 12 hours as soon as
conveniently possible.