Technical information/Data
Effective Composite Curing Wavelength Bands:
40000
35000
30000
25000
20000
15000
10000
5000
Attribute
Information/Specification
Lens
Diameter 9.75 mm
Wavelength range
• Utilizable wavelength range: 385 – 515nm
• Peak wavelengths: 395 – 415nm and 440 – 480nm
Light Intensity Table
Measurement
Instrument
Aperture of Meter
Standard Power
(±10%)
High Power (±10%)
Xtra Power (±10%)
VALO Cordless Curing
Ratings: IEC 60601-1 (Safety), IEC 60601-1-2 (EMC)
Light
Charger Power Supply
Output - 12VDC at 500mA
Input - 100VAC to 240VAC
Ultradent P/N 5930 VALO Charger Power Supply with
international plug inserts
VALO Charger
VALO 3.6VDC Lithium Iron Phosphate smart battery charger:
• Automatic shut off when fully charged
•Auto-detection of defective batteries
•Protections: Thermal, Overcharge, Short-circuit, reverse polarity
o Red LED – Charging
o Green LED – Empty or Fully Charged
o LED off – short circuit
•Charging time: 1 – 3 hours
Rating: CE, WEEE
VALO Batteries
Rechargeable: Safe chemistry Lithium Iron Phosphate (LiFePO4) RCR123A
• Working Voltage: 3.2VDC
Ratings: UL, CE, RoHS, WEEE
Operating Conditions
Temperature: +10°C to +32°C (+50°F to +90°F)
Relative Humidity: 10% to 95%
Ambient Pressure: 700 hPa to 1060 hPa
Duty Cycle:
The curing light is designed for short-term operation. At maximum ambient temperature (32°C) 1 minute ON back-to-back cycling, 30 minutes OFF
(cooling-off period).
8
Effective lower
wavelength band
395 nm - 415 nm
0
360
380
400
420
440
WAVELENGTH (nm)
Nominal Radiant Exitance Comparison Chart
†* Demetron
† MARC
L.E.D.
spectrum
Radiometer
analyzer
7 mm
3.9 mm
1000 mW/cm²
1400 mW/cm²
3200 mW/cm²
(+/-20%)
Effective Upper
wavelength band
440 nm - 480 nm
XTRA POWER MODE
HIGH POWER MODE
STANDARD POWER MODE
460
480
500
520
540
Radiant Exitance will vary based on instrument
capability, measurement method and light
‡ Gigahertz spectrum
placement.
analyzer
† Demetron radiometers and MARC spectrum
analyzers should be used as reference only due to
Exitance
Total Power
having smaller apertures than VALO curing lights.
* Demetron radiometers should be used as
15 mm
15 mm
reference only due to limitations in power and
900 mW/
670 mW
spectral response.
cm²
‡ Radiant Exitance conforms to ISO 10650 when
measured with a Gigahertz spectrum analyzer.
1300 mW/
970 mW
cm²
2100 mW/
1570 mW
cm²
Weight:
• With batteries: 6 oz. (190 grams)
• Without batteries: 5 oz. (150 grams)
Dimension: (8 x 1.28 x 1.06) inches, (203 x 32.5 x 27) mm
Ratings: IEC 60601-1 (Safety)
Cord Length - 6 feet (1.8 meters)
VALO Charger Power Supply is a Medical Grade Class II power supply and provides
isolation from MAINS power