Descargar Imprimir esta página

Medtronic GUARDIAN LINK Manual De Instrucciones página 15

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

  • MEXICANO, página 107
Connecting the tester for testing or cleaning
Before proceeding, have your MiniMed™ insulin pump system user guide available.
To connect the tester:
1
Hold the transmitter and the tester as shown. Line up the flat side of the tester
with the flat side of the transmitter.
2
Push the tester into the transmitter until the flexible side
arms of the tester click into the notches on both sides of the
transmitter.
When properly connected, the green light on the transmitter
flashes 6 times.
3
To test the transmitter, check the sensor icon on the pump
to ensure that the transmitter is sending a signal (see your
system user guide).
4
To clean the transmitter, see Cleaning the transmitter, on page 9.
5
After testing or cleaning, disconnect the tester from the transmitter.
Disconnecting the tester
To disconnect the tester:
1
Hold the transmitter body as shown and pinch the side arms of the tester.
2
With the tester arms pinched, gently pull the transmitter
away from the tester.
Note: To save transmitter battery life, do NOT leave the
tester connected after cleaning or testing.
Cleaning the transmitter
The transmitter is intended for personal use at home (single-
patient use) or for use in healthcare facilities (multiple-patient
use). Single-patient use requires cleaning after each use, while multiple-patient use
requires cleaning and disinfection after each use. When using the transmitter in a
healthcare facility, always follow the cleaning and disinfecting procedure for multiple-
patient use.
WARNING: Do not discard the transmitter in a medical waste container or
expose it to extreme heat. The transmitter contains a battery that
may ignite and result in serious injury.