OBJ_DOKU-193005-006.fm Seite 25 Mittwoch, 31. Oktober 2018 12:40 12
2. Remove surface contaminants with a soft brush/cloth while
holding the product in a water bath and/or rinsing the prod-
uct under running water.
3. Move the movable parts back and forth (at least 5 times).
6.5
Manual cleaning and disinfection
Use a liquid detergent suitable for the preparation of a dipping
bath.
Use a disinfectant compatible with the detergent suitable for the
preparation of a dipping bath.
The detergent and disinfectant must be suitable for medical de-
vices made of plastic and metal and have a pH value between 5.5
and 12.3.
Observe the specifications from the manufacturer of the deter-
gent and disinfectant and in the chapter "Overview of validated
procedures".
1. Prepare a cleaning bath and a separate disinfectant bath.
Cleaning
1. Make sure the instrument jaws are open.
2. Immerse the product completely in a cleaning bath, ensuring
that the product does not come into contact with other com-
ponents in the bath.
3. Thoroughly clean the surfaces with a soft brush/soft dispos-
able cloth.
4. Move the movable parts back and forth (at least 5 times).
5. Rinse the surfaces with sterile, deionized water (at least
1 minute).
6. Repeat the preceding cleaning steps if the product still
shows visible contamination.
Disinfection
1. Make sure the instrument jaws are open.
25