Hach Polymetron 9610sc SiO2 Manual Del Usuario página 12

Ocultar thumbs Ver también para Polymetron 9610sc SiO2:
Tabla de contenido

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

  • MEXICANO, página 42
Option
INSTRUMENT
INFORMATION
CALCULATION
SETUP OUTPUTS
SETUP NETWORK
(conditional)
SECURITY SETUP
AIR PURGE
RESET DEFAULTS
Manage devices
Install or remove input modules.
1. Push menu and select MANAGE DEVICES.
2. Select an option.
Option
SCAN FOR DEVICES The system shows the connected devices.
DELETE DEVICE
Set up the calculation
Set up variables, parameters, units and formulas for the analyzer.
1. Push menu and select CALCULATION.
2. Select an option.
Option
SET VARIABLE X
SET PARAMETER X Selects the parameter referring to the variable X.
SET VARIABLE Y
SET PARAMETER Y Selects the parameter referring to the variable Y.
SET FORMULA
DISPLAY FORMAT
SET UNITS
SET PARAMETER
12 English
Description
Shows the analyzer information. Refer to
on page 18.
Set up variables, parameters, units and formulas for the analyzer. Refer to
the calculation
on page 12.
Selects and configures the 4–20 mA setup, the relay setup and the error hold
mode. Refer to
Configure the outputs
Shows only if a network card is installed. The network card support is for Modbus,
Profibus and HART.
Enables or disables the pass code (default = HACH55).
Enables the air purge for use with an external air supply. Options: on or off
(default). Off: an external air supply is not used. The vent fan is enabled and the
air filter is installed. On: an external air supply is connected to the instrument. The
vent fan is disabled. The air filter is replaced with a fan filter plug. To use this
feature, make sure that the fan filter plug is installed. Refer to the documentation
that is supplied with the air purge kit.
Sets the configuration to the factory defaults.
Description
Note: If no device is connected, the system returns to the main measurement
screen.
Removes the device when the device is no longer connected.
Description
Selects the sensor referring to the variable X.
Selects the sensor referring to the variable Y.
Selects the calculation formula to complete. Options: None, X-Y, X+Y, X/Y, [X/Y]%, [X
+Y]/2, X*Y, [X-Y]%/X
Selects the number of decimal places shown in a calculation result. Options: Auto,
XXXXX, XXXX.X, XXX.XX, XX.XXX, X.XXXX
Enters the unit name (maximum of 5 characters).
Enters the measurement name (maximum of 5 characters).
View instrument information
on page 13 for more information.
Set up

Publicidad

Capítulos

Tabla de contenido
loading

Este manual también es adecuado para:

5500sc sio2Polymetron 9611sc po43-lrPolymetron 9611sc po43-hr

Tabla de contenido