Error
Description
Blood pressure values
Doctor visits can frequently
measured at your doctors
experience anxiety.
offi ce
diff er from your
measurements.
Display remains
Batteries are not installed
blank, or is unusual, when
properly.
instrument is switched on.
Device frequently fails to
Cuff may not be properly
measure blood pressure
positioned.
values, or values measured
are too low (too high).
Further information
Blood pressure is subject to fl uctuations even in healthy people. Comparable
measurements always require the same conditions (quiet conditions)!
In order to receive market clearance from governmental bodies, this device was
subjected to strict clinical tests. The computer program used to measure blood
pressure values was tested by experienced cardiac specialists in Germany.
14
Potential cause and remedy
Record the daily development of the
measured values and consult your
doctor.
Check batteries for
correct polarity.
Check the positioning of the
cuff . See Cuff fi tting on
page 10 for proper fi tting.
Specifications
Model:
BBP2000 / BBP2200
Operation principle:
Oscillometric method
Display:
Liquid crystal display
Measuring range:
20–280 mmHg
Pulse: 40–199 / min
Blood pressure
measurement:
20 mmHg (minimum diastolic value)
280 mmHg (maximum systolic value)
Laboratory accuracy:
± 3 mmHg (cuff pressure)
± 5 % of reading (pulse rate)
Clinical accuracy:
According to AAMI-SP10 with auscultatory reference:
< 5 mmHg systematic off set
< 8 mmHg standard deviation
Infl ation:
Automatic
Power supply:
2 batteries, type AAA 1.5V
Operating temperature /
atmospheric pressure:
+10 °C to +40 °C (50˚F – 104˚F),860-1060 hPa (0.849 - 1.046 atm)
Storage temperature:
–20 °C to +55 °C (-4˚F to +131˚F)
Humidity:
15 to 90 % relative humidity maximum
Cuff :
Suitable for wrist circumference of 13.5 to 21.5 cm
(5.25 - 8.5 inches).
Clinical validation:
According to AAMI-SP10 with auscultatory reference.
(Please note that a validation against intra-arterial
measurements can lead to diff erent results for devices
which are calibrated against auscultatory reference.)
Service Life:
5 years
If device is not used within specifi ed temperature and humidity ranges the technical
accuracy can not be ensured.
Type BF applied parts
See instructions for use
Subject to change without notice.
Operating temperature
55
Storage temperature
15