Mesis FO-3001 Manual De Usuario página 28

Tabla de contenido

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

  • MEXICANO, página 72
Time: time can be adjusted between 0 and 30 minutes. After 30 minutes of
machine operation, it is necessary to wait at least 30 minutes (with the machine
off) before starting a new session.
Caution: if you turn the knob to "Always On" automatic power off is disabled.
Therefore, the machine must be switched off manually.
To increase time, rotate the knob clockwise.
7.2.1 Mode description
Sequential Mode: Chamber ○,1 inflates to pressure set value within maximum 18
seconds. If the chamber pressure reaches preset value within the 18 seconds, holding
pressure on then shift to next chamber inflating. If the chamber pressure doesn't reach
pressure preset value within the first 18 seconds, inflation will be extended for additional
maximum 12 seconds. Within this 12-second period, if the chamber pressure reaches
the preset value, holding pressure on then shift to next chamber inflating. If the chamber
pressure cannot reach the preset value at the end of additional 12 seconds, holding
pressure on then shift to next chamber inflating automatically.
Chamber ○,2 works like Chamber ○,1 . Chamber ○,3 , ○,4 repeats the similar
procedure as Chamber ○,2, ○,3.
After the last Chamber ○,4 completed inflation, Chamber○,1 ○,2 ○,3 ○,4 deflate
simultaneously. The next Chamber ○,1 - ○,2 - ○,3 - ○,4 operation cycle starts 13
seconds after deflation begins.
Ficheiro No.: FSHF-FO3001
28
Version: A/1

Publicidad

Tabla de contenido
loading

Tabla de contenido