OBJ_DOKU-186956-008.fm Seite 12 Donnerstag, 26. November 2020 2:59 14
4.2
Acclimatizing the product
If the product has been stored or transported at temperatures
below +10 °C, allow the product to acclimatize at room tempera-
ture for approx. 3 hours.
4.3
Checking the product
The product, and in particular the connecting cable's insulation,
must be checked for damage before each use.
If damaged, do not use this product!
4.4
Connecting and setting up the product
Connect the footswitch plug into the footswitch socket on the rear
of the unit. Secure the plug in place by tightening the lock nut.
Place the footswitch on a level floor where it cannot tip over.
Do not allow the footswitch cable to stretch above floor level
(risk of tripping). Instead, lay it on the floor.
4.5
Conducting a performance test
Before every use conduct a performance test.
4.6
Footswitch pedals and their function
Footswitch 20188-300:
• Blue pedal: activates coagulation
Footswitch 20188-302:
• Blue pedal: activates coagulation
• Black ReMode button: activates the switch between two set-
ting options in a program
12