Cleaning, disinfection, sterilization
7.1
Safety instructions
Erbe Elektromedizin recommends machine cleaning/
disinfection. Because it is much less effective, manual cleaning
is not recommended.
The following products are suitable for cleaning using an ultra-
sonic bath:
• 20190-045, 20190-047, 20190-062, 20190-065, 20190-
066, 20190-067, 20190-074, 20190-075, 20190-076,
20190-104, 20190-105, 20190-146, 20197-034
The following products are not suitable for cleaning using an ul-
trasonic bath:
• 20190-090, 20190-091, 20190-092, 20190-093, 20190-
094, 20190-095
Never use sharp objects for cleaning.
Disinfectants must be rinsed off thoroughly after use.
Maximum air pressure during drying: 2 bar.
Maximum water pressure during rinsing: 2 bar.
During machine cleaning and disinfection, do not exceed 95 °C.
Observe the specifications from the manufacturer of the deter-
gent and disinfectant and in the chapter "Overview of validated
procedures".
During sterilization, do not exceed 138 °C.
Do not sterilize in hot air.