OBJ_DOKU-180148-009.fm Seite 28 Donnerstag, 17. Januar 2019 10:12 10
7.2
Reprocessing limitation
Frequent reprocessing can damage this product. Observe the
safety instructions for checking the product. With visible damage
or functional impairment, the product may no longer be used.
7.3
Dismantling
1. Remove the HF electrode from the handle.
2. For electrode handle 20190-045 only: Detach the cable from
the handle.
7.4
Required aids
Reprocessing step Items Needed
Precleaning
Soft plastic brush/soft cloth
Manual cleaning/
Soft plastic brush/soft disposable cloth
disinfection
(low particle)
Compressed air gun (2 bar maximum)
7.5
Precleaning
Use water for precleaning, a non-fixing disinfectant if necessary.
1. Remove surface contaminants with a soft brush/cloth while
holding the product in a water bath and/or rinsing the prod-
uct under running water.