Equipment; Recommended Separation Distances - EMS AIR-FLOW MASTER Manual De Instrucciones

Ocultar thumbs Ver también para AIR-FLOW MASTER:
Tabla de contenido

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

  • MEXICANO, página 22

recommended separation distances

The AIR-FLOW MASTER is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The customer or the user of
the AIR-FLOW MASTER can help to prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications
equipment (transmitters) and the AIR-FLOW MASTER as recommended below, according to the maximum output power of the communications equipment.
rated maximum output power of
transmitter
W
0.01
0.1
1
10
100
For transmitters rated at a maximum output power not listed above, the recommended separation distance d in metres (m) can be estimated using the
equation applicable to the frequency of the transmitter, where P is the maximum output power rating of the transmitter in watts (W) according to the trans-
mitter manufacturer.
NOTE 1:
At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2:
These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects and
people.
8 8
FB-348_3_ed_2008-06_EMC AIR-FLOW MASTER.indd 8
separation distance according to frequency of transmitter
150 kHz to 80 MHz
80 MHz to 800 MHz
d = 1.2 √P
0.12 m
0.38 m
1.2 m
3.8 m
12 m
m
800 MHz to 2.5 GHz
d = 1.2 √P
0.12 m
0.38 m
1.2 m
3.8 m
12 m
d = 2.3 √P
0.23 m
0.73 m
2.3 m
7.3 m
23 m
02/06/08 12:50:33