Calibration Menu - OHAUS Pioneer PX323/E Manual De Instrucciones

Ocultar thumbs Ver también para Pioneer PX323/E:

Publicidad

Idiomas disponibles

Idiomas disponibles

Pioneer
The mode menu has the following sub-menus:
.
.MODE
COUNT
SET ON
SET OFF
PERCENT
SET ON
SET OFF
DYN
SET ON
SET OFF
DENSITY
SET ON
SET OFF
END.MOD
Note: * Use No key to change the temperature value
Enter Menu – When the balance is on, press and hold Unit/Menu until [mMENU] appears.
Release the button and the Calibrate [.CAL.] menu will display.
Menu Navigation – Select menus, menu items and settings through use of the Yes and No buttons.
Solid arrows point to the content displayed when Yes is pressed, Dashed lines when No is pressed.
Changing Settings – To select the displayed Setting, press Yes. To move to the next Setting, press No.
Exit Menu – When [.END.] is displayed, press Yes to exit the menu function, or press No to return to the Cal menu.
Note: Press and hold No at any time to exit quickly.

4.2 Calibration Menu

InCal or Span calibration should be performed daily and when the room temperature changes.
InCal [InCal] calibrates the balance using an internal mass.
Span Calibration [SpaN] uses two weight values: zero and a weight between 50% and 100% of the capacity of the
balance.
Linearity calibration [LIN] uses three weight values zero, 50% of capacity and full capacity. Generally this
calibration is not required unless testing shows that the linearity error exceeds the Linearity tolerance in the
Specification table. (Not available in InCal models)
Calibration Adjust [CAL Adj] allows adjustment to the result of the internal calibration by +/- 99 divisions.
(InCal models only)
.setup.
4.3 Setup Menu
[
Automatic Zero-Setting [azsmM] – Environmental changes can cause the display to drift. The Automatic Zero-
Setting Mechanism (AZSM) is designed to keep the balance set at zero, despite these slight changes. (OFF,
SET.5d, SET 1d, SET 2d, SET 5d)
Filter [filter] – Use the low setting (SET LO) when environmental disturbances are not present. Use the
medium setting (SET MED) for normal environments. Use the high setting (SET HI) when vibrations or air currents
are present.
Auto Tare [A-tare] – The initial item placed on the balance is assumed to be a container so it is zeroed out. The
next item is then weighed. When the pan is cleared the balance resets, waiting for a container. (SET OFF, SET
ON)
APWOPT
SET ON
SET OFF
AUTO
SET ON
SET OFF
SOLID
WATER
OTHER
LIQUID
SINKER
.CAL.
[
]
]
T20.0C*
01.00000g/cc
010.000g/cm
EN-6

Publicidad