Elcometer 138 Guia Del Usuario página 14

Ocultar thumbs Ver también para 138:
Tabla de contenido

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

  • MEXICANO, página 62
4 USING THE CONDUCTIVITY METER (continued)
2
Dip the meter into a solution of known temperature and allow
time for the temperature reading to stabilise.
3
Press 'INC' or 'DEC' to enter calibration mode. The 'CAL'
indicator appears on the display, immediately press '°C/°F' to
switch to temperature calibration mode.
When you enter calibration mode, if the conductivity reading is outside the
specified range 'Er.1' is displayed. You can still proceed to temperature
calibration mode by pressing '°C/°F' immediately. If '°C/°F' is not pressed
within two seconds, the meter exits calibration mode and returns to
measurement mode.
4
The upper display shows the measured temperature reading
based on the last set offset (if any) and the lower display shows
the default (uncalibrated) temperature reading based on factory
settings.
5
Use the 'INC' and 'DEC' buttons to adjust the upper temperature
reading to the known temperature value of the solution.
The temperature adjustment window is ±5°C (±9°F) from the default
.
reading
6
Wait for 5 seconds for the meter to automatically conrm the
temperature calibration by displaying 'CO' and returning to
measurement mode.
'Er.0' is displayed and the meter returns to measurement mode if the
temperature of the solution is not within 0°C to 50°C (32°F to 122°F).
Press 'HOLD ENT' to exit temperature calibration mode without confirming
the calibration.
Dipped in a
solution of known
temperature
en-13
R
Figure 7: Temperature Calibration Sequence
INC
DEC
Calibration Mode
°C/°F
Temperature Calibration Mode
Measurement
Mode
INC
DEC
Conrmation
displayed for 2
seconds
www.elcometer.com

Publicidad

Tabla de contenido
loading

Tabla de contenido