1. Press the red button A on the control unit B to stop
the boom and patient from raising or lowering.
2. To reset, rotate the emergency button clockwise.
5.5 Activating an Emergency Release
Emergency release with CBJ Home control unit
If the hand control fails, the boom can be lowered by using
the circular switch for emergency release.
1. Lower the boom by pressing and holding button A at
the front of the control unit.
2. Stop lowering the boom by releasing the button.
1575480-E
Emergency Release with Jumbo Care control unit
C
A
1. Insert a pen into the hole labeled Emergency Up A or
Emergency Down B on the control unit C.
Activating an emergency release manually
(Not available on Birdie® Compact)
If the emergency lowering on the control unit is not working,
a mechanical emergency lowering is available as a back-up.
This may happen in case of partial or total power failure, or
if the battery runs down during use.
Usage
B
19