Symbols
Full Battery
Low Battery
PUL. DISPLAY ERROR
SYS.
DIA.
PUL./min.
AM
PM
In Canada - Auto Control Medical (Toll-Free): 1-800-461-0991
Function / Meaning
Appears during measurement when
the cuff is applied loosely.
Previous measurements stored in memory
Average data
The battery power indicator during
measurement.
The battery is low when it blinks.
Unstable blood pressure due to
movement during measurement.
The systolic and diastolic values are
within 10 mmHg of each other.
The pressure value did not increase
during the inflation.
The cuff is not applied correctly.
The pulse is not detected correctly.
Blood pressure monitor internal error
Systolic blood pressure in mmHg
Diastolic blood pressure in mmHg
Pulse per minute
Data taken in the AM
Data taken in the PM
Pressure settings
Indicates the pressure value as set
by the user.
WHAT DISPLAY SYMBOLS MEAN
Recommended Action
The reading may yield an
incorrect value. Apply the
cuff correctly, and take
another measurement.
Replace all batteries with new
ones when the mark blinks.
Take another measurement.
Remain still during
measurement.
Apply the cuff correctly,
and take another
measurement.
Remove the batteries and
press the START button,
and then install the batteries
again. If the error still
appears, contact the A&D.
E-8