• RV 10.5 Vertical-condenser with manifold and cut-off valve
for reflux distillation
Also available with coating (RV 10.50)
• RV 10.6 Vertical-intensive condenser with manifold and
cut-off valve for reflux distillation
This vertical-intensive action condenser features a double jacket
design for particularly efficient reflux distillation.
Also available with coating (RV 10.60)
Hose system
• Connect the water hoses (H
the counter-flow principle. (Fig. 20)
• Install the vacuum connections to the condenser, Woulff bottle,
vacuum controller with valve and vacuum pump.
• Always connect the vacuum hose (vac) to the condenser at the
highest point to minimize solvent losses during suctioning.
• Use standard laboratory vacuum hoses with an inside diameter
of 8 mm and a wall thickness of 5 mm (see accessories).
28
O) to the condenser according to
2
Fig. 20
Heating bath
Refer to the chapter entitled "Operation"
CAUTION
in the operating instructions of heating
bath!
• Assemble the foot of the heating bath
For operation of the heating bath in combination with the RV
8, the foot plate supplied must be fitted to the heating bath.
Caution: It is not permitted to use heating bath fitted with
a foot plate in combination with any other type of rotary
evaporator.
• Place the heating bath on the stand of the rotation drive and
push it into the left position. (Fig. 22)
Fig. 21
Fig. 22