Disposal
7
Disposal
Dispose of the Dräger X-act 5000 Basic according to local waste
disposal regulations.
7.1
Disposal in Europe
This product must not be disposed of as household waste. It
is therefore marked with the adjacent symbol. You can return
this product to Dräger free of charge. For information please
contact the national marketing organizations or Dräger.
Batteries must not be disposed of as household waste. They
are therefore marked with the adjacent symbol. Dispose of
batteries at battery collection centers as specified by the
applicable regulations.
8
Technical data
Ambient conditions
Temperature during storage
Temperature during operation 5 to 40 °C (41 to 104 °F)
Instrument temperature for
measurements at low ambient
temperatures
Relative humidity
Pressure
Dimensions (H x W x D)
Weight (without power supply) approx. 1.6 kg
20
-20 to 55 °C (-4 to 131 °F)
≥5 °C (≥41 °F)
0 to 95 % r.h., not condensing
700 to 1300 hPa
(10.2 to 18.9 psi)
approx. 175 x 230 x 108 mm
(approx. 7" x 9" x 4.25")
(approx. 3.5 lb)
Power supply
Rechargeable battery
(NiMH Battery)
Short-term measurements
Sample taking
Flow range
Flow
Resolution
0.1 to 1.0 L/min
0.1 L/min
1.0 to 2.0 L/min
0.2 L/min
1)
If the flow rate for a single sample taking measurement is additionally
adjusted, then the tolerance is equal to the tolerance of the used flow meter.
If, for example, the used flow meter has a tolerance of ±3 %, then the
adjusted flow rate of the Dräger X-act 5000 Basic has a tolerance of ±3 %.
7.2 V, 1500 mAh
Charging current: 400 mA,
Max. charging voltage: 15 V
Charging time: <4 hours
Adjustable,
1 to 200 strokes
depending on the used tube
Stroke volume: 100 mL ± 5 mL
Sampling time
adjustable, up to 12 hours
depending on the adjusted flow rate
Resolution:
in steps of 15 minutes (default setting)
in steps of 1 minute
(can be set in the password-protected
menu)
0.1 to 2.0 L/min
Tolerance
1)
±10 %
1)
±10 %
Dräger X-act 5000 Basic