General information

MF-Model is a vaccine storage freezer and
icepack freezer. It has two compartments;
one for vaccine storage and one for icepack
freezing.

Power supply

The MF vaccine freezer is equipped with
a heavy-duty compressor that can operate
at volta ges 22% below nominal voltage.
See the rating plate. If voltage variations go
beyond this range it is recommended to use
an automatic voltage stabilizer.

Technical data

This device complies with relevant EU
directives including Low Voltage Directive
2006/95 EEC. and Electromagnetic Compa-
tibility Directive 2004/108/EC
The rating plate provides various techni-
cal information as well as type and serial
number.
IMPORTANT! There must be free access to
the ventilation grille
Unpacking and
installation
Unpacking
1. Unpack the appliance and check that it
has not been damaged. If you observe
any damage, inform your supervisor.
2. Check the rating for correct voltage and
frequency.
3. Open and remove all internal packing
material.
Installation
The appliance must be placed in a well ven-
tilated room, not in direct sunlight and away
from other heating sources.
Place the appliance on the floor, ensure that
it is level. Distance between the wall and the
back and left side of the appliance must be
minimum 50 mm and the distance be tween
the wall and the right side of the appliance
must be minimum 100 mm. (fig. 2).
If installed in a room with other refrigerators/
freezers, ensure a minimum distance bet-
Fig. 2
ween the appliances of minimum 200 mm.
There must be at least 30 mm clearance
between the base of the appliance and the
floor.
5
Only MF 214 and MF 314
GB