INSTRUCTIONS
A) Site Selection
• Chose a site on the chest where the skin is clean of debris and dry prior to sensor placement.
• Ensure orientation and location of the sensor on the patient matches the sensor site location in the Masimo device
settings.
Chest
Refer to Fig. 1. The preferred site is on the patients chest.
B) Applying the sensor
1. Refer to Fig. 2. Open the package and remove the sensor.
Note: Do not remove the release liner at this point.
2. Refer to Fig. 3. Pull to remove the plastic battery tab.
3. Refer to Fig. 4. Move the sensor close to the Masimo device (such as Root®), to enable Bluetooth pairing.
Note: Refer to the Masimo device Operators Manual for complete instructions.
4. Clean and dry the sensor application site.
5. Refer to Fig. 5. Pull off the release liner from the sensor.
6. Apply the sensor on to the selected application site.
Note: Centroid sensor is designed for removal and reapplication for no more than two (2) times over the life of the product.
C) Removing the Sensor
1. Peel gently to remove the sensor from the patient.
Note: Disposal of Product: Comply with local laws in the disposal of the Centroid sensor, host device and/or its accessories.
SPECIFICATIONS
The sensors have the following specifications:
Centroid Sensor:
Weight Range
Application Site
Sensor Life
Respiratory Rate Range*
Respiratory Rate Accuracy
*A
accuracy is a statistical calculation of the difference between device measurements and reference measurements. Approximately two-thirds of the
rms
device measurements fell within +/- A
Respiration rate performance has been validated against manual scored capnogram respiratory measurements on 40 healthy volunteer subjects and
1
34 hospitalized adults. The clinical testing results may not be generalized to all patient conditions.
ENVIRONMENTAL
Centroid Sensor:
Storage/Transport Temperature
Shelf Life
Operating Temperature
Storage/Transport Humidity
Operating Humidity
Protection against harm from liquid ingress

WIRELESS SPECIFICATIONS

Communication (Bluetooth)
Type
CAUTION: In order to maintain Bluetooth connectivity with the device, ensure that the Centroid sensor is within specified
distance and line of sight of the host device. See the host device Operator's Manual.
COMPATIBILITY
This sensor is intended for use only with devices containing Masimo technology. Each sensor is designed to
operate correctly only on the systems from the original device manufacturer. Use of this sensor with other devices
may result in no or improper performance.
For Compatibility Information Reference: www.Masimo.com
approximately 30 g
Chest
96 hours
8 – 35 breaths/min
3 rpm
1
of the reference measurements in a controlled study.
rms
-20°C - 50°C @ ambient humidity
2 years
10°C - 40°C @ ambient humidity
15% RH - 90% RH (non-condensing) @ ambient temperature
15% RH - 95% RH (non-condensing) @ ambient temperature
IP24, Protection from ingress of particulates and water spray from any direction
Configurable Settings
Bluetooth Low Energy
4
9836D-eIFU-0421