SPEcIFIcAtIONS

catalog Number
Delivery Rate (Lower delivery rates
available for low flow applications)
maximum Recommended Flow (@
nominal outlet pressures of zero and 7
kPa)
Outlet Pressure
Electrical Rating
Operating Voltage Range
Oxygen Percentage
Operating Altitude
(tested at 21°C only)
0-1500 M (0-4921 ft)
1500-3000 M (4921-9842 ft)
3000-4000 M (9842-13123 ft)
Operating Environment Range*
10°C to 35°C, humidity range of 30%
to 75%
Power Consumption
Weight
Sound level (ISO 8359:1996 from
front)
Dimensions
Pressure Relief Valve
SE-515X
DEvILbISS 4-LItER AND 5-LItER SERIES
515ADS/515ADZ
515DS/515DZ
1 to 5 LPM
5 LPM
8.5 psig (58.6 kPa)
515A Series - 115 V, 60 Hz, 4.1 Amp
515 Series - 115 V, 60 Hz, 4.2 Amp
97-127 V~, 60 Hz
1-5 LPM=93%±3%
Across the voltage range:
No degradation of performance
515A Series
Tested at nominal voltage only:
No degradation in performance expected
based on tests of similar products.
515 Series
Tested at nominal voltage only:
No degradation in performance
Not recommended/Not tested
515A Series - No degradation in
performance up to 40°C across the
operating voltage range. (tested at 670 m).
515 Series - No degradation in performance
across the operating voltage range (tested
@ sea level)
515A Series - 385 Watts Average
515ADS Only - 320 Watts @ 2.5 LPM &
below
515 Series - 400 Watts Average
515A Series - 50 lbs. (22.7 Kilograms)
515 Series - 52 lbs. (23.5 Kilograms)
515A Series - 48 dbA overall average
515 Series - 52.5 dbA overall average**
27.75"H x 16"W x 14"D
(70.5 x 40.6 x 35.6 cm)
515A Series - 40 psig±5psig
(276 kPa±34.5 kPa)
515 Series - 44 psig±3psig
(303 kPa±21 kPa)
515AKS/515AKZ
515KS/515KZ/515NS
1 to 5 LPM
5 LPM
8.5 psig (58.6 kPa)
515A Series -220-230 V~, 50 Hz, 1.7 Amp
230 V~, 60 Hz, 2.4 Amp
515 Series - 220-230 V~, 50 Hz, 1.7 Amp
230 V~, 60 Hz, 1.9 Amp
187-255 V~, 50 Hz
218-253 V~, 60 Hz
1-5 LPM=93%±3%
Across the voltage range:
No degradation of performance
515A Series
Tested at 230V/50Hz only:
No degradation in performance expected
based on tests of similar products.
515 Series
Tested at 230V/50Hz only:
~4% oxygen performance drop at 5 LPM
Not recommended/Not tested
515A Series - No degradation in
performance up to 40°C across the
operating voltage range. (tested at 670 m).
515 Series - ~4% oxygen performance drop
at 5 LPM and 35˚C condition. No
degradation in performance across the
operating voltage range at other liter flows
(tested at 670 M).
515A Series - 230V / 50 Hz - 330 Watts
Average
230V / 50 Hz - 305 Watts Average 515AKS
Only @ 2.5 LPM & below
515 Series - 230V / 50 Hz – 365 Watts
Average
515A Series - 230V / 60 Hz - 450 Watts
Average
230V / 60 Hz 515AKS Only - 415 Watts
Average @ 2.5 LPM & below
515 Series – 230V / 60Hz – 435 Watts
Average
515A Series - 50 lbs. (22.7 Kilograms)
515 Series - 53 lbs. (24.5 Kilograms)
515A Series - 48 dbA (50Hz) overall
average
515 Series - 50.5 dbA (50Hz) overall
average
27.75"H x 16"W x 14"D
(70.5 x 40.6 x 35.6 cm)
515A Series - 40 psig±5psig
(276 kPa±34.5 kPa)
515 Series - 44 psig±3psig
(303 kPa±21 kPa)
515UK 4-Liter
1 to 4 LPM
4 LPM
8.5 psig (58.6 kPa)
230 V~, 50 Hz, 1.4 Amp
195-253 V~, 50 Hz
1-4 LPM=93%±3%
Across the voltage range:
No degradation of performance
Not recommended/Not tested
Not recommended/Not tested
No degradation in performance across
the operating voltage range (tested @
sea level).
295 Watts Average
53 lbs. (24.5 Kilograms)
50.5 dbA overall average
27.75"H x 16"W x 14"D
(70.5 x 40.6 x 35.6 cm)
44 psig±3psig (303 kPa±2.1 kPa)