6 CALIBRATING AND TESTING THE GAUGE CALIBRATION
To calibrate (using the glass zero tile):
1
Press the Cal softkey
2
Select "Calibrate" (Model T)
3
Follow the on-screen instructions.
To test calibration (Model T):
1
Press the Cal softkey
2
Select "Test Calibration" and follow the on-screen instructions
3
When prompted, either place the probe on to the glass zero tile or on the centre of the foil
resting on the glass tile, making sure the pin goes through the hole in the centre of the foil.
4
After continued use, the precision manufactured probe tip will wear, increasing the tip size and
will then need to be replaced. This is a user replaceable item. Please contact your Elcometer
sales representative to order new probe tips and tip replacement tool.
7 BATCHING (MODEL T)
1
To use the Batching memory facility, press the Batch softkey
2
Select "New Batch" or "Open Existing Batch" to add readings
3
Copy and review batch data
4
Select "Edit Batch" to rename, clear readings from or delete a batch
5
Fixed Batch Size allows users to pre-define the number of readings to be stored in a batch.
Once all readings have been taken the gauge automatically opens a new batch with a link
to the original batch name. For Example NewBatch_1 becomes NewBatch_2, NewBatch_3, etc.
4
Save each reading into memory or store the average of a pre-defined number of readings
using the Counted Average function.