Cleaning Instructions - Roscoe Medical ASP-ROS Manual De Instrucciones

Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

4. Place the patient tubing to the appropriate position of patient and start to
suck sputum aspirate.
5. When the sputum reaches the safe full level of collection (800 cc), the
suctioning will stop sucking automatically (float shut off).
6. Switch the power switch to "O" position (Off), shut off the suction motor.
7. Remove the sputum from collection bottle after the motor stops turning
completely. (Refer to the "Cleaning" section below for the instructions on how
to clean the collection bottle.)
8. Turn on the unit and start to run the suction unit again.
NOTE
CAUTION
WARNING
CLEAN COLLECTION BOTTLE
1. Turn off the aspirator and allow vacuum to drop, then disconnect AC power
source.
2. With the collection bottle still in the holder, remove the collection bottle lid.
The bottle can now be taken out of the holder to be emptied.
NOTE
3. The collection bottle should be thoroughly cleaned after each use by one of
the following three methods:
a.
Wash in a hot water/dishwashing detergent solution and rinse with
clean, hot tap water. Then wash in one part vinegar to three parts hot
water solution. Rinse with hot tap water and air dry.
b.
Wash with rubbing alcohol and air dry.
1. If the unit sustains a severe drop, validity of the gauge
must be checked. Inspect if the float shut-off is still
functioning properly when the liquid in the collection
reaches the safe full level.
2. Always transport unit with the vacuum adjust knob
rotated fully clockwise in case unit is dropped.
1. When automatic float shut-off is activated, contents of the
collection bottle should be emptied. Further suctioning
could cause damage to the vacuum pump.
2. Should fluid be aspirated back into the unit, discontinue
use immediately and send your aspirator to an authorized
service center for a detailed inspection as the vacuum
pump may have been damaged.
If the device has been operating continuously for 30
minutes or more, turn off power and allow the device to
cool for at least 30 minutes before restarting.
CLEANING
Collection bottle should be emptied after each use.