STERILIZATION:
Warning:
Prior to sterilizing the clip applier instrument, make sure cartridge is removed.
Microline PENTAX shall not be responsible for any damages occurring as a result of a
non-compliance with this warning.
1.
Prior to sterilization, the instrument must be thoroughly cleaned.
2.
Wrap the instrument or use sterilization tray (Cat #3705/3707).
3.
Sterilize the instrument following the protocol provided by the sterilizer manufacturer.
Microline PENTAX recommends the following as minimum sterilization cycle requirements:
Gravity Cycle:
Pre-Vacuum Cycle:
INSPECTION:
NOTE:
M/L-10 Inventory Inspection Procedure (to be performed once a month)
An abbreviated inspection should be performed in the OR each time before handing the M/L-10 to
the surgeon for use:
1. Check jaws for misalignment, damage and caked on bio- burden.
–
2. Check shaft for straightness. Remove outer tube and check inner shaft for damage and
–
caked on bio-burden.
3. Squeeze and release handle 20 plus times to check for sticking. Check handle for damage.
–
Never pry handle and trigger open.
4. Fire a minimum of 2 clips to check clip closure, engagement into jaws and clip/jaw
–
alignment.
5. In addition to proper and thorough cleaning (see instructions) make sure the M/L-10 is
–
properly lubricated between cases:
» A. Fully immerse the instrument in an approved instrument lubricant.
» B. If unable to fully immerse the instrument make sure to thoroughly lubricate both jaw
and rotation knob area.
» C. After lubrication process perform step 3 and 4 of the inspection if the applier handle
was sticking or locking.
WARRANTY:
Microline PENTAX warrants that this instrument is free from defects in both material and
workmanship.
Microline PENTAX shall not be held liable for incidental or consequential damage of any kind. This
warranty is valid only to the original purchaser of the instrument and for a period of one (1) year.
Work performed on this instrument by anyone other than a Microline PENTAX Authorized Service
Center will void this warranty.
SERVICE AND REPAIR:
Prior to returning this instrument for repair, call your Microline PENTAX Distributor to obtain authorization.
The instrument will be returned un-repaired to the sender if the following conditions are not met:
•
The instrument must be given a "Return Goods Authorization" (R.G.A) number.
•
The instrument must be cleaned and sterilized prior to returning for repair.
•
The R.G.A. Number must be clearly visible on the outside of the box it is shipped in.
P/N: 09-39-30 Rev D ECO 949 04/06
3 minutes @ 270F (132C)
3 minutes @ 270F (132C)
-7-
ENGLISH